Introduction
Visual field is a measurable part of visual function. Interpretation of visual field is an integral part of evaluation of many ophthalmic conditions such as glaucoma, neuro-ophthalmology and retinal pathologies. Standard automated perimetry (SAP) is the most widely used method to assess visual field deficit in glaucoma.1 Many articles have been published on perimetry interpretation, but none of them have covered about the recent advances in perimetry. In this article, we will learn about the recent advances and their significance in the current era of modern perimetry.
Algorithms- Evolution
Currently in perimetry, adaptive mode of threshold estimation is used in which the stimulus luminance varies in ascending or descending steps until the threshold is estimated. This process is also known as staircase or bracketing. Early algorithms initiate examination from four principal stimulus locations called anchor points (Octopus) or seed points (Humphrey Field Analyser) followed by 4-2-1 dB bracketing strategy which crosses the threshold twice. Second generation algorithms include Dynamic strategy, wherein the step size increases from 2dB to 10dB in Octopus, when the visual field defect increases. FASTPAC algorithm of Humphrey applied 3dB steps either in ascending or descending fashion,but the threshold is crossed only once. Current algorithms used are Swedish Interactive Thresholding Algorithm(SITA), Zippy Adaptive Threshold Algorithm(ZATA), Tendency Oriented Perimetry(TOP), German Adaptive Thresholding Estimation(GATE-i/GATE), Adaptive Staircase Thresholding Algorithm (ASTA), Continuous Light Increment Perimetry(CLIP), SPARK Precision and SPARK Quick.
ZATA was introduced for Henson 8000 perimeter with two versions, Standard and Fast. It makes use of retinal sensitivity if available from previous tests to set the starting intensity for each test location thus saving considerable time. TOP was first introduced in 1996 for Octopus perimeters, it uses linear interpolation between test locations. TOP uses the response of subject at a specified point to estimate the sensitivity at that point and also modify sensitivity approximation of surrounding points, thus decreasing the slope of the boundary around visual field deficits.2 GATE-i/GATE is similar to Full Threshold staircase and the SITA Standard strategy it starts by determining the sensitivity at five predefined seed locations. The threshold is defined as the mean of the dimmest seen stimulus and the brightest not seen stimulus. The ASTA Standard algorithm for baseline examinations uses a 4-2-2 staircase approach as well as neighbouring test target information and a quick termination methodology to reduce test time. ASTA Fast is a short test protocol advised for patients expected to be “within normal limits” or those who have previously shown unusual levels of fatigue while ASTA Follow-Up further reduces the test time during follow-up examinations. ASTA alogorithms are used in Heidelberg edge perimetry. CLIP is a fast threshold strategy, using stimuli with constantly rising luminance, offered for use with the Oculus Easyfield perimeter. In this algorithm stimulus luminance is continuously increased in smaller steps(usually 1dB), from an infrathreshold level according to the patient’s reaction time until it is seen. The CLIP algorithm has been found suitable for the examination of children above the age of 8 years. SPARK strategy was first introduced on Oculus Easyfield perimeter. The SPARK Quick strategy is good for follow-ups and for screening examinations, it also offers a separate training strategy to reduce the curve of the learning effects in standard perimetry. SITA standard is analogous to Full Threshold algorithm, while SITA fast is analogous to FASTPAC. In 2019 Heijl and associates introduced a new time saving threshold visual field testing strategy called SITA Faster which is intended to replace SITA Fast.3
Advances In Testing Strategies
SITA Faster: At least five threshold visual field testing is required to quantify how quickly the patient is losing visual field sensitivity and there is a need of much frequent visual field testing in a newly diagnosed glaucoma patient.4 But in actual practice, perimetric testing is considerably lower than recommended due to lack of resources.5 In order to achieve higher number of testing, SITA faster strategy was introduced which is 50% faster than SITA standard testing and uses the same SITA algorithm and normative data as SITA fast.(Figure 1)
SITA Faster starts much closer to the expected threshold in the principal points and necessitates only 1 reversal at prime test points as compared to two staircases in earlier SITA tests, also perimetrically blind points are not rechecked. Due to shorter duration of the test, patients with severe glaucoma appreciated SITA Faster the most. The testing time is also reduced by not using blind spot and false negatives detection in analysis; instead it uses gaze monitoring and false positives for test quality monitoring. SITA faster also has the advantage for GPA analysis and regression analysis wherein the software allows comparative interpretation of the three SITA thresholding strategies because of strong similarity between their visual field index findings.6
Pre-Perimetric Tests
As much as 35-50 % ganglion cells can be lost before a visual field defect is detected, thereby emphasing the importance of techniques to diagnose early glaucoma.7 A brief of the genesis behind such tests is enumerated in Table 1.
Short wavelength automated perimetry (SWAP) also known as blue on yellow perimetry is a specialised technique in which a blue type V stimulus is used on a yellow background of 100 candela per square metre. Introduced in 1993, SWAP is based on the principle that yellow background light reduces the responsiveness of red and green cones by saturating them, so as to mainly test the blue cones, which are primarily affected in early glaucoma patients.8 It helps to detect development of glaucoma in ocular hypertension (OHT) patients and progression of disease in early stages. In a study done by VK et al, SWAP was found to have a very high specificity compared with most of the Spectral Domain OCT(SD-OCT) parameters in differentiating glaucomatous optic neuropathy from non-glaucomatous eyes.9 However SWAP has high test-retest variability and is highly affected by cataract and media opacities, thus it is not solely used for glaucoma management.
Frequency Doubling Technology (FDT)
FDT presents an achromatic sinusoidal grating stimulus of low spatial frequency undergoing counterphased flickering at a high temporal frequency, leading to apparent doubling of the spatial frequency of the grating, known as frequency doubling illusion.10 FDT offers many advantages such as high sensitivity and specificity for diagnosing early glaucomatous damage11, not greatly affected by refractive error and cataract.
FDT stimulus predominaths stimulates magnocellular ganglion cell pathway(My cells) involved in motion and flicker detection (Table 1). FDT has two presentation patterns, C-20 which represents the central 20 degree with 17 stimulus locations and N-30 which has two additional points in the nasal field. There are screening and threshold tests both of which compare individual test results with the normative database of more than 700 eyes of 450 subjects.12 The C-20-1 has got higher specificity, while the N-30-5 has got higher sensitivity for detecting glaucomatous field damage. The second generation matrix FDT has improved spatial resolution of visual field defects and can detect small localized defects due to smaller target size and increased number of testing locations. Matrix threshold tests use Zippy Estimation of Sequential Thresholds (ZEST) algorithm which is based on Bayesian Statistics.13 Matrix FDT has many advantages over the first generation as it has a video eye monitor for patient alignment tracking with an option to pause the test in between, patient display screen is larger thus preventing any movements to change fixation for seeing nasal points, data storage option and enhanced statistical analysis for better evaluation and interpretation of test results (Figure 2).

High Pass Resolution Perimetry (HRP)
HRP assesses the function of “P ganglion cells” which are concentrated in the central retina. The stimulus consists of a range of ring targets with dark borders and lighter centres stimuli (14 sizes) (Figure 3) used at 50 locations, the sensitivity is verified by varying the size and not the luminance of the stimulus. The stimulus distribution of the HRP corresponds with the arrangement of the ganglion cells, therefore it could be superior to SAP in detection of visual field defects.14 The HRP demonstrates less variability at visual field locations with reduced sensitivity than SAP, however studies comparing HRP with SAP shows variable results.15,16
Rarebit perimetry (RBP)
This type of perimetry uses stimulus nearer in size to an individual ganglion cell (rarebits, microdots)17 to identify abnormal function of any one fixed retinal ganglion cell. (Figure 4) The test uses 0,1 and 2 suprathreshold dots at 24 rectangular test areas and checks for the patient response. By evaluating a number of combinations of dots in small localized regions, it determines visual performance (detection or “hit” rate) in these areas. Nearly all rarebits are seen with a normal retina (close to 100% "hit rate"), while disorders of the visual system result in losses from missing or dysfunctional receptive fields which appear as gaps in the receptive field matrix. The outcome of the algorithm was adversely affected by optical defocus and by cataract.18 RBP has been used as a screening method to detect central vision defects such as macular lesions, optic neuropathies and chiasmal lesions.
Motion Perimetry
Motion detection is largely detected by large ganglion cells via magnocellular pathway19 and a damage to this pathway is associated with loss of motion perception.20 Glaucoma patients who have been tested with full-field motion stimuli demonstrate deficits in motion perception.21 The test background consists of randomly arranged white dots on a grey background, while the motion targets are circle random dot cinematograms randomly placed in order to reduce the effect of positional cues.22 (Figure 5) These stimuli are of 17 different sizes with a diameter step factor of 100.1(1.259) and a 2-1 staircase is used to estimate threshold. The test evaluates 44 locations and progresses until the smallest circle is seen at each test point, making it time- consuming.
Flicker defined form (FDF)
This test consists of flickering random black and white dots on a background of 50 cdm2 of mean luminance, creating an imaginary edge outline due to phase differences between the stimulus and the background (stimulates the magnocellular pathway)23 (Figure 6) The Heidelberg edge perimeter (HEP) is a monitor-based perimeter that uses the FDF stimulus to test central and peripheral visual fields. It offers a full set of standard tests for central 10°, 24° and 30° visual field as well as an extension to the periphery up to 60° (100 points) and uses the Adaptive Staircase Thresholding Algorithm (ASTA) for visual sensitivity estimation. There is a considerable learning effect over three visits for the Edge stimulus, therefore there is a lack of agreement between the HFA and FDF.
Pulsar Perimetry
This technique can evaluate both parvocellular and the magnocellular visual pathway.24 It consists of two images the phase and counterphase image that alternate with a frequency of 10 Hz over 500 ms and merge with the background luminance of 32 cd/m2 at the edges to avoid stimulating direction-selective ganglion cells. The Pulsar examination method of the Octopus 600 exclusively uses the Tendency Oriented Perimetry in which correlation of the threshold values in neighbouring locations is taken into account thereby reducing the examination time by nearly 80%. The variability was lower for Pulsar, compared to both SAP and FDT.25 Pulsar perimetry seems able to detect more cases of clear progressive glaucomatous damage than either confocal scanning laser ophthalmoscopy or nerve fibre polarimetry.26
Flicker Perimetry
This perimetry stimulates M ganglion cells. It consists of three different techniques: Temporal Modulation perimetry (TMP), Luminance Pedestal Flicker perimetry (LPF) and Critical Fusion Frequency perimetry (CFF). TMP reveals greater defects in early glaucoma at all temporal frequencies and thus identifies cases of ocular hypertension that are likely to develop glaucoma.27 LPF perimetry is commercially available in Medmont M600 perimeter and demonstrates a flickering stimulus, superimposed on a base of a steady luminance and specifies the temporal frequency required to separate the stimulus from the base.28 CFF perimetry determines the highest temporal frequency at which a flickering stimulus of constant luminance is originally perceived as a continuous(non-flickering) stimulus.29 A few studies have reported flicker perimetry superior to SAP, in the investigation of glaucomatous field loss.30
Microperimetry/ Fundus Perimetry
It is a novel functional method which assesses retinal sensitivity while directly examining ocular fundus, thus correlating the pathology(nerve fibre layer loss, macular pathology etc.) and the corresponding functional loss.31 The fundus is imaged in real-time and the visual field is directly mapped on it, thereby providing a direct structural and functional correlation. An eye tracker detects eye movements during stimulus projection, allowing accurate matching between expected and projected stimulus on the retina. Currently, there are three popular models: NIDEK MP3, Optos OCT/SLO, MAIA microperimeter (Centervue). The Compass perimeter (Centervue Padova, Italy) is a combination of scanning ophthalmoscope and automated perimeter, providing confocal images of the retina with retinal threshold sensitivity under non-mydriatic conditions. It has strategies similar to 24-2 and 10-2 programmes of SAP, along with recent addition of ZEST algorithm. There are three different testing modes namely Fast test, Expert test and Follow up test. The Fast test is done for uncooperative patients requiring only 3 minutes per eye. Microperimeter can be used for diagnosis and progression in early glaucoma cases, and evaluation of retinal sensitivity in localized areas of retinal nerve fibre layer defects (pre-perimteric glaucoma analysis)( Figure 7).

Advances in testing patterns 24-2c
The standard 24-2 algorithm of perimetry is known to miss the early macular damage which has been identified on 10-2 and OCT parameters.32 The Humphrey 24-2C visual field test is a modified 24-2 visual field test that incorporates 10 additional test points in the central 10° of vision to aid in the detection of this missed damage. It was introduced on the Humphrey platform (HFA3) in 2019. The additional 10 points in the center are asymmetric in the two hemifields and are chosen as per studies indicating the more susceptible areas for early ganglion cell damage (Figure 8). It is available in the SITA faster testing mode by default, which helps keep the testing time comparable to 24-2 despite the addition of 10 extra points. This novel testing pattern helps to study the structure-function concordance in early glaucoma and suspects. Comparative studies have been done evaluating its role in glaucoma as well as neuro-ophthalmology.33 24-2c testing pattern has been compared with the standard 24-2 and 10-2 patterns. The results of comparative studies so far have been variable, citing no additional benefit to detection of extra field defect areas, although structural-functional concordance was not accurate.34 The 24-2C exhibits the potential to be used as a hybrid between the 24-2 and 10-2 to better evaluate visual field defects.35

Advances in Analytical indices
Glaucoma Hemifield Test (GHT)
GHT is an artificial intelligence-based analysis that provides plain language classifications of 24-2 and 30-2 test results (points common to both) and is based upon patterns of loss specific to glaucoma (not neurological disorders).36 Pattern Deviation scores in each of the five zones in the upper hemifield are compared to findings in mirror-imaged zones in the inferior visual field and compared to normative significance limits specific to each zone pair (Figure 9).
GHT findings are divided into the following categories:
i. Outside Normal Limits- One zone pair differs by an amount found in fewer than 1 % of normal subjects.
ii. Borderline- At least one zone pair differs by an amount found in 1% to 3% of normal subjects.
iii. General Depression Best test point locations have low sensitivity to the level seen in fewer than half a percent of normal subjects.
iv. Abnormally High Sensitivity- Best test points are so high as to be at levels seen in fewer than half a percent of normal subjects.
v. Within Normal Limits.
The GHT has been reported to have high sensitivity and specificity37, with easy interpretation, especially by novice users. The method was designed to have an overall specificity between 84% to 90% depending on whether the Borderline findings are considered Outside Normal Limits or Within Normal Limits.
Visual Field Index (VFI)
Introduced by Bengtsson and Heijl in 2008, the Visual Field Index is an improved version of the older mean deriation (MD) index, that is less affected by cataract than MD, except in fields having MDs worse than -20dB. As opposed to MD, VFI is expressed in the percentage of normal age-corrected visual function, and is intended for use in assessing glaucoma progression and staging glaucomatous functional damage on a scale from normal function (100%) to perimetric blindness(0%). MD value associated with blind fields depends upon age and testing strategy, while VFI in a perimetrically blind field is always 0%, regardless of age or strategy and 100% in normal fields. VFI gives more weightage to the central test points as compared to the peripheral ones to account for the higher density of ganglion cells found in the central retina.38
Guided Progression Analysis (GPA)
Over the past decade, a paradigm shift has occurred in glaucoma management from monitoring progressive perimetric glaucomatous damage on serial field testing to current focus on the patient’s rate of disease progression, relying heavily on the quantification and velocity of visual field loss over time. Defining a baseline test is important, as it could be the beginning of glaucoma therapy or a significant modification in the glaucoma treatment plan. GPA has been programmed to choose by default the earliest two fields as baseline tests.
GPA can be event based (Glaucoma Change Probability Maps) and trend based (VFI), both of which are reflected in the standardized GPA reports (Figure 10). The event based analysis determines visual field progression to be either present or absent, based on a predefined change in the parameters, and gives a point wise glaucoma change probability analysis. The progression analysis shows a numerical plot and a plot using intuitive symbols indicating progression of the disease. A solid triangle represents a point changing by an amount that is significant (P<0.05) and is repeated in three consecutive follow-up exams; half filed triangle identifies a point changing by an amount that is significant (P<0.05) and is repeated in two consecutive follow-up exams; small open triangle symbol is used when the change was not seen on the previous follow up test. A significant deterioration in the probability maps at same three or more points on two subsequent field tests predict a possible progression, whereas that on three subsequent tests determine a likely progression. The trend-based analysis is computed by the visual field index which provides the actual rate of change of visual field parameters and is based on linear regression analysis. The ‘p’ value of VFI slope determines the change of VFI with time and a negative slope represents disease progression. The VFI Bar histogram graphically represents the patient’s current VFI value along with a 3-5 year projection of the VFI regression line, considering the same rate of progression is maintained. Mixed GPA program now allows intermixing of different strategies such as SITA Faster, Fast, Standard, 24-2, 30-2, and 24-2C in progression analysis due to their clinical equivalence.

Advances in Data Integration
Data synchronization and review is now available in recent models which keeps all perimeters connected through a local network, thereby decreasing patient waiting time, as fields may be performed on any one of the connected perimeters, with constant synchronisation of patient’s fields over all connected perimeters.
Integrated Glaucoma Workplace has also been introduced in the advance models of Humphrey perimeters to enable digital synchronisation and integration of various glaucoma imaging modalities such as Humphrey visual field perimeter, OCT-RNFL, Ganglion cell analysis and macular GCC. This will enable an elaborate structural-functional correlation of the disease pathology and progression.
Advances in patient alignment and fixation monitoring
The initial perimeters utilised manual assessment of patient alignment by an extrenal examiner during the duration of the test., however this type of assessment was gross and had a lot of demerits including inter-examiner variability. In 1974, Heijl-Krakau method was first described in which moderately bright stimuli are presented at the expected location of the physiological blind spot of the tested eye and any positive response to the stimuli suggested a loss of fixation.39 The method though still widely used in many automated perimeters, has a few shortcomings because the patient’s blind spot is not always situated in the location assumed, thus making the fixation loss stimuli a positive response during the examination. Also, the amount of gaze data provided by this method is quite limited, because only a few blind spot check stimuli can be presented during a visual field test.
The corneal view in the machine screen provides useful data regarding patient's pupil and lens centration using a green star, red star and green plus coding (Figure 11).
The Gaze tracker is a novel patient alignment monitoring system present in Humphrey perimeters and measures gaze direction every time a stimulus is presented. The patient’s alignment is displayed in the form of lines, marks with a precision of approximately +/- 2 degrees and is presented at the bottom of the HFA printout. A line with full-scale marking extending upward indicates an off fixation i.e gaze error of 10 degrees, a small downward mark indicates that the gaze tracker was unable to measure gaze direction, while larger downward mark indicate eyelid interference with the device’s view of the eye such as in blinking or squinting (Figure 12).
The HFA gaze tracker uses image analysis to separately locate the centre of the pupil and the reflection of the light-emitting diode from the corneal surface. The spacing between these two points strongly depends upon gaze direction while being largely independent of the head position. Separate calculations provide head position information and are used to automatically keep the eye aligned at the centre of the trial lens.
Binocular Testing Strategy
Esterman Binocular Field Testing
This binocular field testing method available on the HFA utilises size III white stimuli of 10dB intensity to test 120 points in the central and peripheral visual field and records the visualisation response at each point. Esterman test can be used for driver’s license screening and is as effective as HVF in detection of central defects in case of advanced glaucoma.40 Binocular testing allows for naturally-occuring binocular enhancement wherein either eye compensates for the defects in the fellow eye, thereby making it more functional and relevant for the patient. However, it is difficult to judge whether the defect is absolute or relative, and there is no way to control fixation stability since the binocular testing conditions eliminate naturally occurring blind spots which are used for fixation control in other tests.
Crabb et al in 1998 described another new method known as Integrated Visual Field (IVF) which simulates a binocular visual field from the data of monocular visual fields of either eye.41 The study revealed a substantial agreement between the simulated binocular results and the binocular Esterman test in classifying glaucomatous patients.
Advances in Refractive Error Correction
Liquid Trial Lens
Refractive error correction prior to initiating visual field testing is crucial for obtaining optimal test results. One diopter of spherical refractive blur in an undilated pupil will produce approximately one decibel of depression of the hill of vision when testing with a Goldmann size III stimulus. This can be avoided by performing perimetry with the patient wearing his refractive correction in the form of glasses or contact lenses (along with near add in presbyopic age group) or by manually placing the corrective trial lenses in the perimeter holder. A newly introduced liquid trial lens technology has proven to be equally effective, convenient and time saving technology for refractive correction. It automatically adjusts to the spherical and presbyopic power of the patient by changing the fluid pressure within the lens which changes its shape from concave to convex (Figure13). The power adjustment can be made between +8.0 D to -8.0 D range. It is recommended to use spherical equivalent corrections for astigmatism upto 2.0 D. The user interface contains + and – buttons to change the liquid trial lens power in 0.25 D steps, while the patient is asked whether the target is perceived as blurred or focussed before the procedure.

Advances in Perimetric Unit
Portable Perimeters
The idea of portable visual field testing was borne in order to facilitate monitoring of presumed stable or controlled glaucoma patients or suspects reducing the overall number of office visits for the patient. Although the portable perimeters have a low resolution and are not as accurate as compared to the in-office gold standard method, aggregating more data over a longer period of time may be more predictive than the smaller amount of higher quality data that’s acquired only during annual or semi-annual office visits. Portable perimeters can be used in the office setup as well as for home screening and follow-up. Three most popular portable perimeter designs at present are Moorfields Motion Displacement Test(MDT), Melbourne Rapid Fields Test and Virtual Reality Peripheral Vision testing. MDT was one of the early portable visual field test in which 32 white lines on a gray background was used while fixating in the centre and any wiggling movement was documented by the patient with a button click. However, MDT is no longer in use and not available. Melbourne Rapid Field Test uses a free Ipad software and takes advantage of a moving fixation target in order to effectively increase the tablet surface area to test up to 30 degrees of field. Several published studies have validated the device as a screening tool similar to a tangent perimeter.42 Virtual Reality peripheral Vision testing overcomes one of the major limitations of portable devices i.e. inability to control fixation and the screen to eye distance. With the virtual reality, no matter where the patient looks, the stimuli can be shown relative to fixation at that moment. Gyroscopes can account for head movement, and immersive environment can improve user engagement. Various inexpensive, light weight, mobile VR applications and software platforms are now available. Tsapakis et al showed a high correlation between the realibility of VF testing using a VR testing system and Humphrey test.43 However, further long term studies are needed to validate portable perimeters.
Conclusion
Perimetry has come a long way from the era of kinetic perimeters consisting of heavy equipments to modern automated static perimeters consisting of faster and portable computers. The future of perimetry is now moving towards artificial intelligence for automated analysis of visual fields for the detection of glaucoma and predicting future progression. Another aspect of research and development is a portable brain-computer interface that detects visual function through recorded electric responses directly from the visual cortex thereby eliminating the subjective aspect of clicking to indicate when a visual stimulus is seen.44
References
- Advanced Glaucoma Intervention Study. 2 . Visual field test scoring and reliability. Ophthalmology. 1994 Aug;101(8):144555.
- Morales J, Weitzman ML, Gonzalez de la Rosa M. Comparison between Tendency- Oriented Perimetry (TOP) and octopus threshold perimetry. Ophthalmology. 2000 Jan;107(1):134-42.
- Heijl A, Patella VM, Chong LX, Iwase A, Leung CK, Tuulonen A, et al. A new SITA perimetric threshold testing algorithm: Construction and a multicenter clinical study. Am J Ophthalmol. 2019 Feb;198: 154-65.
- Chauhan BC, Garway-Heath DF, Goni FJ, Rossetti L, Bengtsson B, Vishwanathan AC, et al. Practical recommendations for measuring rates of visual field change in glaucoma. Br J Ophthalmol. 2008 Apr;92(4):569-73.
- Fung SS, Lemer C, Russell RA, Malik R, Crabb DP. Are practical recommendations practiced? A national multi-centre cross-sectional study on frequency of visual field testing in glaucoma. Br J Ophthalmol. 2013 Apr;97(7):843-84.
- Heijl A, Bengtsson B, Patella VM. Glaucoma follow-up when converting from long to short perimetric threshold tests. Arch Ophthalmol. 2000 Apr;118(4):489-93.
- Sit AJ, Medeiros FA, Weinreb RN. Short-wavelength automated perimetry can predict glaucomatous standard visual field loss by ten years. Semin Ophthalmol 2004 Sep;19:122-4.
- Kanski JJ, Bowling B. Clinical Ophthalmology: A Systematic Approach. 7th ed. Philadelphia: Elsevier/Saunders; 2011. P.3348.
- Kalyani VK, Bharucha KM, Goyal N, Deshpande MM. Comparison of diagnostic ability of standard automated perimetry, short wavelength automated perimetry, retinal nerve fiber layer thickness analysis and ganglion cell layer thickness analysis in early detection of glaucoma. Indian J Ophthalmol. 2021 May;69:1108-12.
- Kelly DH. Frequency doubling in visual responses. J Opt Soc Am. 1966;56:1628-33.
- Cello KE, Nelson-Quigg JM, Johnson CA. Frequency doubling technology perimetry for detection of glaucomatous visual field loss. Am J Ophthalmol. 2000 Mar;129:314-22.
- Adams CW, Bullimore MA, Wall M, Fingeret M, Johnson CA. Normal ageing effects for frequency doubling technology perimetry. Optom Vis Sci. 1999 Aug;76(8):582-7.
- Turpin A, McKendrick AM, Johnson CA, Vingrys AJ. Properties of perimetric threshold estimates from full threshold, ZEST and SITA-like strategies, as determined by computer simulation. Invest Ophthalmol Vis Sci. 2003 Nov;44(11):4787- 95.
- Spry PG, Johnson CA, Mansberger SL , Cioffi GA. Psychophysical investigation of ganglion cell loss in early glaucoma. J Glaucoma. 2005 Feb;14(1):11-9.
- Ennis FA, Johnson CA. Are high-pass resolution perimetry thresholds sampling limited or optically limited? Optom Vis Sci. 2002 Aug;79(8):506-11.
- Artes PH, Chauhan BC. Longitudinal changes in the visual field and optic disc in glaucoma. Prog Retin Eye Res. 2005 May;24(3):333-54.
- Hackett DA, Anderson AJ. Determining mechanisms of visual loss in glaucoma using Rarebit perimetry. Optom Vis Sci. 2011 Jan;88(1):48-55.
- Nilsson MO, Abdiu CG, Laurell, Martin L. Rarebit perimetry and fovea test before and after cataract surgery. Acta Ophthalmol. 2010 Jun;88(4):479-82.
- Merigan WH, Byrne CE, Maunsell JH. Does primate motion perception depend on the magnocellular pathway? J Neurosci. 1991 Nov;11:3422-9.
- Schiller PH, Logothetis NK, Charles ER. Functions of the colour-opponent and broad-band channels of the visual system. Nature 1990;343:68-70.
- Bullimore MA, Wood JM, Swenson K. Motion perception in glaucoma. Invest Ophthalmol Vis Sci. 1993 Dec;34:3526-33.
- Nakayama K, Tyler CW. Psychophysical isolation of movement sensitivity by removal of familiar position cues. Vision Res. 1981;21:427-33.
- Horn FK, Scharch V, Mardin CY, Lammer R, Kremers J. Comparison of frequency doubling and flicker defined form perimetry in early glaucoma. Graefes Arch Clin Exp Ophthalmol. 2016 May;254(5):937–46.
- Gonzalez-Hernandez M, Garcia-Feijoo J, Sanchez Mendez M, Gonzalez de la Rosa M. Combined spatial, contrast and temporal functions perimetry in mild glaucoma and ocular hypertension. Eur J Ophthalmol. 2004 Nov-Dec;14(6):514–22.
- Gonzalez de la Rosa G, Hernandez G, Estevez A, Aleman D, Plasencia A. Diagnostic capability of pulsar, FDT and HRT-II in glaucoma suspects. Arch Soc Esp Oftalmol. 2007 Jul;82(7):413-22.
- Gonzalez de la Rosa M, Gonzalez- Hernandez M, Diaz-Aleman T. Linear regression analysis of the cumulative defect curve by sectors and other criteria of glaucomatous visual field progression. Eur J Ophthalmol. 2009 May-Jun;19(3):416–24.
- Wang X, Xu K, Zhang H, Zhao J, Zhu X, Wang Y, Wu R. Retinal ganglion cell death is triggered by paraptosis via reactive oxygen species production: A brief literature review presenting a novel hypothesis in glaucoma pathology. Molecular Medicine Reports. 2014;10: 1179-83.
- Bernardi L, Costa VP. Shiroma LO. Flicker perimetry in healthy subjects: Influence of age and gender, learning effect and short-term fluctuation. Arq Bras Oftalmol. 2007 Jan-Feb;70(1):91-9.
- Roberti G, Manni G, Riva I, Hollo G, Quaranta L, Agnifili L, et al. Detection of central visual field defects in early glaucomatous eyes: Comparison of humphrey and octopus perimetry. PLoS One. 2017 Oct;12(10):e0186793
- Matsumoto C, Takada S, Okuyama S, Arimura E, Hashimoto S, Shimomura Y. Automated flicker perimetry in glaucoma using Octopus 311: A comparative study with the Humphrey matrix. Acta Ophthalmol Scand. 2006 Apr;84:210-5.
- Ratra V, Ratra D, Gupta M, Vaitheeswaran K. Comparison between Humphrey Field Analyzer and Micro Perimeter 1 in normal and glaucoma subjects. Oman J Ophthalmol 2012 May;5:97-102.
- Grillo LM, Wang DL, Ramachandran R, Ehrlich AC, De Moraes CG, Ritch R,et al. The 24-2 Visual Field Test Misses Central Macular Damage Confirmed by the 10-2 Visual Field Test and Optical Coherence Tomography. Transl Vis Sci Technol.2016 Apr 14;5(2):15.
- Yamane MLM, Odel JG. Introducing the 24-2C Visual Field Test in Neuro-Ophthalmology. J Neuroophthalmol. 2021 Dec 1;41(4):e606-e611. doi: 10.1097/WNO.0000000000001157. PMID: 33417411.
- Phu J, Kalloniatis M. Ability of 24-2C and 24-2 Grids to Identify Central Visual Field Defects and Structure-Function Concordance in Glaucoma and Suspects. Am J Ophthalmology.2020 Nov:219:317-31.
- Phu J, Kalloniatis M. Comparison of 10-2 and 24-2C Test Grids for Identifying Central Visual Field Defects in Glaucoma and Suspect Patients. Ophthalmology. 2021 Oct;128(10):1405-16.
- Asman P, Heijl A. Evaluation of methods for automated Hemifield analysis in perimetry. Arch Ophthalmol. 1992 Jun;110(6):820-6.
- Katz J, Sommer A, Gaasterland DE, Anderson DR. Comparison of analytic algorithms for detecting glaucomatous visual field loss. Arch Ophthalmol. 1991 Dec;109(12):1684-9.
- Bengtsson B, Heijl A. A visual field index for calculation of glaucoma rate of progression. Am J Ophthalmol. 2008 Feb;145(2):343-53
- Heijl A, Krakau CE. An automatic static perimeter, design and pilot study. Acta Ophthalmol. 1975 Jun;53(3):293-310.
- Ayala M. Comparison of the monocular Humphrey visual field and the binocular Humphrey esterman visual field test for driver licensing in glaucoma subjects in Sweden. BMC Ophthalmol 2012 Aug 2;12:35.
- Crabb DP, Viswanathan AC, McNaught AI, Poinoosawmy D, Fitzke FW, Hitchings RA: Simulating binocular visual field status in glaucoma. Br J Ophthalmol 1998 Nov; 82(11):1236–41.
- Johnson CA, Thapa S, George Kong YX, Robin AL. Performance of an iPad application to detect moderate and advanced visual field loss in Nepal. Am J Ophthalmol 2017 Oct;182:147-154.
- Tsapakis S, Papaconstantinou D, Diagourtas A, Droutsas K, Andreanos K, Moschos MM, et al. Visual field examination method using virtual reality glasses compared with the Humphrey perimeter. Clin Ophthalmol 2017 Aug;11:1431-43.
- Nakanishi M, Wang YT, Jung TP, Zao JK, Chien YY, Diniz-Filho A, Daga FB, et al. Detecting glaucoma with a portable brain-computer interface for objective assessment of visual function loss. JAMA Ophthalmol 2017 Jun;135(6):550-7.